Specification Monitor HI-981520-02
pH
Range 0.0 to 14.0 pH - (depending on electrode capabilities - see individual specifications)
Resolution 0.1 pH
Accuracy ±0.2 pH
Calibration Manual or automatic two-point calibration in 7.01 and 10.01 pH
ppt
Range 0.0 to 70.0 ppt (g/L)
Resolution 0.1 ppt (g/L)
Accuracy ±1.0 ppt between 0.0 ppt and 40.0 ppt
±2.0 ppt between 40.0 ppt and 70.0 ppt
Calibration Automatic, single point in 35.00 ppt
PSU
Range 0.0 to 70.0 PSU
Resolution 0.1 PSU
Accuracy ±1.0 PSU between 0.0 and 40.0 PSU
±2.0 PSU between 40.0 and 70.0 PSU
S.G.
Range 1.000 to 1.041 S.G.
Resolution 0.001 S.G.
Accuracy ±0.001 S.G.
Temperature
Range 0.0 to 50.0 °C
Resolution 0.1°C
Accuracy ±0.5°C
Compensation Automatic, 5 to 50°C
Alarm
pH High or Low with enable or disable options
EC High or Low with enable or disable options
Temperature High or Low with enable or disable options
High/Low value with High/Low alarm enabled
Probes pH - HI-1286; EC - attached
Power supply 12 vdc adapter (included) from 115 Vac, and 230 Vac
Environment 0 to 50°C; RH max 95%, non-condensing
Casing IP65 ingress protection
Dimensions 125 x 185 x 38mm
Weight 300g
Ordering information HI-981520 Marine monitor is supplied with HI-1286 pH electrode; attached EC and Temperature probe; pH7.01 buffer solution 20ml x2; pH 10.01 buffer solution 20mL x 2; 35.00 ppt salinity calibration solution 20mL x 2; electrode cleaning solution 20mL x 2; suction cap with clip x 2; self-adhesive fasteners x 2; 12 Vdc power adapter; quick reference guide with QR code for manual download and instrument quality certificate.